Description:
Steam automatic forevacuum sterilizer with passage (double door type) is simple and easy in use and environmental friendly.
Model GPD-560-1is equipped with import components.
It is intended for pressurized water-saturated steam sterilization of medical application devices made of metal, glass, rubber, ligature suture material and textile, steam action on which does not affect their functional properties. The sterilizer is used in patient care and other medical institutions. It is installed between “clean” and “sterile” zones and perfect for using in Sterilization Departments. For easy transportation and installation the sterilizer’s design gives possibility to be detached in 2 blocks: chamber and power unit*.
Notes: To prevent from scale formation and, therefore, to increase efficiency of the sterilizer’s operation it is required to use distilled water.
Technical data:
Volume of chamber, dm3
560
Dimensions of chamber (L x W x H), mm, no more than
1050õ630õ900
* Overall Dimensions (L x W x H), mm
1670õ1340(980õ360)õ1800
Loading height, mm
750
1st Program (toC - min.- MPa)
134 – 5 –
0,21
2nd Program (toC - min.- MPa)
121 – 20
– 0,11
Free programmable mode (toC - min.- MPa)
120÷135 -
5÷180 – 0,1÷0,21
Capacity of built-in steam generator , kg/h of steam
50
Voltage, V
380
Rated power, kVA
45
Weight, kg, max.
1520
Advantages: · Simple and easy in use, environmental friendly; · The main units and panels are made of high-alloy stainless steel; · Sterilization chamberis manufactured from high-alloy stainless steel grade 12Õ18Í10Ò (AISI 321); · The chamber is rectangular-shaped; · Built-in steam generator; · Automatic microprocessor control, operation by means of «one button»; · Possibility of programming of the sterilization cycles; · Powerful vacuum pump provides effective air removal from sterilization chamber by method of pulsing evacuation and vacuum drying of the sterilized products; · Vacuum at drying is at least -0,08 MPa; · Residual humidity of sterile textile products is less 1%; · Cyrillic font printer to make documentation of sterilization parameters; · Electronic video-graphic recorder is used as an independent device for making documentation; · LCD in Cyrillic font for visual indication of sterilization parameters; · Light and sound indication of sterilization stages, completing of cycles, malfunctions; · Re-usable bacterial filter to clean air supplying to the ster. chamber at the stage of pressure equalizing; · Vacuum-test (for testing tightness of all systems) and Bowie-Dick test (for testing efficiency of air replacing with steam); · Nonvolatile clock, calendar, cycle counter; · Air-operated valves are used; · The sterilizer is equipped with built-in steam suppressor having automatic control of the temperature of condensate draining to the sewerage system; · Service life is 10 years; · Upon the Customer request it is possible to complete with loading devices: loading trolleys with containers, big and small baskets.
Safety System: · Interlock of starting of sterilization cycle, if door is open; · Door locking during the program running; · Door locking in case of chamber is under pressure; · Interlock of the program running at emergency situations. · Electric heaters disconnection in case of decreasing of water level in steam generator; · Protection devices from overpressure in steam generator; · Password protection to avoid unauthorized changing of the parameters; · For diagnostics and testing the control system can be connected to PC including remote PC via modem; · Manual control at emergency situations.
Standard complete set: · Spare parts kit – 1 set. · Shelves – 3 pcs. · Remote manual control desk – 1 pc. · Compressor AIRBAG CV.
During installation of the sterilizer it is recommended to purchase panel strips for making full contact of the sterilizer’s finish and room surfaces. The panel strips give possibility to maintain pressure difference between “clean” (“non-sterile”) and “sterile” zones.